ISMED, NOVRIZAL (2015) APLIKASI SENSOR TEKANAN MPX5100DP PADA TENSIMETER DIGITAL BERBASIS MIKROKONTROLLER ATMEGA 8535. Other thesis, POLITEKNIK NEGERI SRIWIJAYA.
|
Text (Cover)
COVER.pdf Download (176kB) | Preview |
|
|
Text (BAB I Pendahuluan)
BAB I.pdf Download (140kB) | Preview |
|
|
Text (BAB II Tinjauan Pustaka)
BAB II.pdf Download (630kB) | Preview |
|
![]() |
Text (BAB III Rancang Bangun Alat)
BAB III.pdf Restricted to Repository staff only Download (314kB) |
|
![]() |
Text (BAB III Flowchart)
Flow Chart BAB III.pdf Restricted to Repository staff only Download (93kB) |
|
![]() |
Text (BAB IV Pembahasan)
BAB IV.pdf Restricted to Repository staff only Download (606kB) |
|
|
Text (BAB V Kesimpulan dan Saran)
BAB V.pdf Download (21kB) | Preview |
|
|
Text (Daftar Pustaka)
DAFTAR PUSTAKA.pdf Download (264kB) | Preview |
Abstract
This final report explains how to work and the application of pressure sensor detects MPX5100DP Systolic and Diastolic blood pressure in the microcontroller ATMega8535 as media-processing data. Processed data will be displayed on the LCD microcontroller. From the resulting data, the writers can do the calculation value of the blood pressure measurement sensors MPX5100DP. Pressure sensors on digital tensimeter this, generates an analog output voltage which has a linear value to the value of the pressure readings on the sensor. The yield on the sensor will be processed by ADC on a microcontroller that can be generated analog to digital conversion as a result of sensor readings. When the tool is run, the dc motor will live to carry out the air filling the cuff tensimeter, this is done so that the sensor can detect MPX5100DP Systolic and Diastolic blood pressure. After reading the pressure value by the sensor, its results will be processed by a microcontroller which will be displayed on the LCD screen. From the measurement results using a digital tensimeter MPX5100DP sensor is compared with a digital tensimeter manufactured. Note the differences that occur in the percentage of 1% - 5%. This difference occurs because MPX5100DP detection sensor is too sensitive, so it is necessary to enhance tensimeter additional sensor measurements that the piezoelectric vibration sensor vibration wave filter of 1Hz so that a more accurate reading.
Item Type: | Thesis (Other) |
---|---|
Uncontrolled Keywords: | MPX5100DP sensor , microcontroller ATMega8535 , Systolic , Diastolic |
Subjects: | T Technology > TK Electrical engineering. Electronics Nuclear engineering |
Divisions: | Electronic Engineering > Undergraduate Theses |
Depositing User: | Mrs Trisni Handayani |
Date Deposited: | 07 Apr 2016 02:20 |
Last Modified: | 07 Apr 2016 02:20 |
URI: | http://eprints.polsri.ac.id/id/eprint/2809 |